cve: "CVE-2019-6538"
severity: "CRITICAL"
cvss: 9.3
epss: "0.9%"
vendor: "Medtronic"
kev: false
exploited: false
published: "2019-03-25 22:29:00"
tags: [cve, security, critical]
source: tsecurity.de CVE-Dossier
exported: "2026-10-03T11:45:31+02:00"
---

# CVE-2019-6538

> 9.3 CRITICAL

## Beschreibung

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

## CVSS-Vektor

```
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
```

| Metrik | Wert | Bewertung |
|---|---|---|
| AV Angriffsvektor | Benachbart | warn |
| AC Komplexität | Gering | bad |
| PR Privilegien | Keine | bad |
| UI Interaktion | Keine | bad |
| S Scope | Verändert | bad |
| C Vertraulichkeit | Keine | good |
| I Integrität | Hoch | bad |
| A Verfügbarkeit | Hoch | bad |

## Referenzen

- <https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01>
- <http://www.securityfocus.com/bid/107544>

---
_Exportiert aus dem [tsecurity.de CVE-Dossier](https://tsecurity.de/cve?cve=CVE-2019-6538) · Datenquellen: EUVD (ENISA), NVD, OSV, CISA KEV, FIRST EPSS, Exploit-DB, BSI WID (CSAF)_